| Code | Text | Description |
|---|---|---|
|
-1600 |
Power off requested |
Robot power had been turned off by request. One of the following occurred:\
|
|
-1601 |
Software Reset: using default settings |
When the system was restarted, the "Software Reset" switch on the MCIM Selector Switch was set to the ON state. This forced the system to read in the default configuration files (*.PAC) instead of the standard files. This feature should be used if a configuration files becomes corrupted or a setting inadvertently make the system unusable. |
|
-1602 |
External E-STOP |
A front panel E-Stop loop ("ESTOP_L 1" or "ESTOP_L 2") has been broken and the status signal "External ESTOP_L" is asserted, indicating that external equipment is the source of the E-Stop. |
|
-1603 |
Watchdog timer expired |
The hardware watchdog timer on the CPU board has expired and robot power is disabled. This error may indicate a hardware failure or software bug and should not normally be seen. If this error persists, please contact customer service to report this problem. |
|
-1604 |
Power light failure |
The robot power-on light interfaced via the front panel connector has burned out. Robot power is turned off and may not be turned back on. Please contact customer service. |
|
-1605 |
Unknown power off request |
Robot power is off or was turned off, but no request was made to turn it off. This error may indicate a hardware failure or software bug and should not normally be seen. |
|
-1606 |
E-STOP stuck off |
When the controller is restarted, a diagnostic program has detected an E-Stop circuit that is stuck in the off state. Robot power cannot be turned on until the stuck E-Stop circuit is repaired and the controller is restarted. |
|
-1607 |
Trajectory task overrun |
The periodic trajectory generation system task was unable to complete its executing during its allotted period. The parameter "Trajectory Generator update period in sec" (DataID 600) is set too small for the type of robot you are using. |
|
-1609 |
E-STOP timer failed |
When the controller is restarted, a diagnostic program has detected that the hardware E-Stop timer is not triggering as expected. Robot power cannot be turned on and you must restart the controller. If this error persists, please contact customer service to report this problem. |
|
-1610 |
Controller overheating |
In GPL 3.0 H1 and later, this error
is reported in addition to -1617 CPU overheating, -3144 Amplifier
overheating or -3145 Motor overheating. These other
errors provide detailed information on the specific component
that is generating the overheating error condition. |
|
-1611 |
Auto/Manual switch set to Manual |
A attempt has been made to enable robot power, using one of the standard means, when the Auto/Manual switch is set to Manual. The robot power has not been enabled. The standard means of enabling power are by using: the PowerEnabled property of the Controller class in GPL, the "Enable Power" parameter (DataID 241), the "Automatic power on" parameter (DataID 240), the power enable digital input (defined by DataID 242), or CANopen. |
|
-1612 |
Power supply relay stuck |
An attempt to turn on power has failed because an internal diagnostic test indicates that the motor power supply relay is stuck in the "on" position and may be unsafe. Reboot your controller. If this error persists, contact customer service. |
|
-1613 |
Power supply shorted |
Power has been disabled because the motor power supply has detected that it is shorted. Check the wiring and the amplifiers to make sure no short is present. |
|
-1614 |
Power supply overloaded |
Power has been disabled because the motor power supply has detected an overload condition. Verify that you are not attempting to draw more than the rated current from the power supply. Verify that the parameter "Delay after turning on power in sec" (DataID 263) is set long enough to allow power to stabilize before the amplifiers are enabled. |
|
-1615 |
No 3-phase power |
A power supply has been configured to use 3-phase power, but the 3rd phase is not connected or has failed. It is still possible to run the power supply in this mode at low power, but attempting to draw high power may overheat the power supply. |
|
-1616 |
Shutdown due to overheating |
The CPU exceeded its operating temperature for too long a period of time and the controller automatically turned off. When the CPU first exceeds its maximum permitted limit of 90 degrees Celsius, a "Controller overheating" error (-1610) is generated and motor power is disabled. If the temperature does not decline within 5 minutes, error code -1616 is logged to the flash in a special system file ("/flash/system/errlog.txt")and the controller is shutdown. The next time the controller is restarted, the -1616 error code will be displayed in the error log. This error code will be displayed every time the controller is restarted until the error log is cleared. |
|
-1617 |
CPU overheating |
Either the master or a slave CPU board temperature has exceeded its maximum temperature of 90 degrees Celsius. See CPU temperature (DataID 126) for the current temperature of all CPUs. If the temperature does not fall below 90 degrees, the controller will switch off in 5 minutes and the controller must be restarted. Normally this error is followed by the generic error -1610 "Controller overheating". Ensure there is good air circulation around the controller and check that the fans are not blocked. The Installation section of the controller hardware manuals provide guidelines for ventilating and cooling controllers. |
|
-1618 |
Power supply not communicating |
Some PreciseFlex™ power supplies communicate with the PreciseFlex™ controller for identification, error detection, and safety interlocks. This communications has failed. Be sure that all cables between the controller and power supply are installed properly. Verify that DataID 128 "Power supply type" is set properly for the type of power supply you are using. |
|
-1619 |
Power disabled by GIO timeout |
This errors occurs in two possible cases:
|
|
-1620 |
Safety diagnostics failed |
In enhanced CAT3 mode, the safety diagnostic checks have failed. Robot power cannot be enabled. Previous error messages may indicate the cause of the failure. Use the "Show Safety" web console command for more details. Once the problem has been resolved, try to enable power again. |
|
-1621 |
Safety software configuration mismatch |
The safety configuration bits in DataID 2031 "Enhanced safety mode" do not match the settings of DataID 117 "Safety mode" or require hardware not present in your controller. For example, you set bit &H001 of DataID 2031 but DataID 117 is not set to 4 or 5 or you set bit &H200 in DataID 2031 but do not have a 3-phase power supply. Verify that DataID 117 is correct. For hardware-related errors, contact Brooks support. If a numeric value is displayed with this error, the value indicates what aspect of the safety configuration did not match. See DataID 2031 "Enhanced safety mode" for a description of the bits that form this value. |
|
-1622 |
Not allowed in safety mode |
You have attempted to perform some operation that is not allowed in your current safety mode. For example, you have attempted to enable power from a user program while DataID 117 "Safety mode" is set to 5 (Enhanced CAT-3 full). |
|
-1623 |
ESTOP1 stuck on |
The safety diagnostics in enhanced CAT3 mode have determined that ESTOP channel 1 is asserted. Probably your channel 1 ESTOP loop is open. Close the loop and retry the operation. |
|
-1624 |
ESTOP2 stuck on |
The safety diagnostics in enhanced CAT3 mode have determined that ESTOP channel 2 is asserted. Probably your channel 2 ESTOP loop is open. Close the loop and retry the operation. |
|
-1625 |
ESTOP1 stuck off |
The safety diagnostics in enhanced CAT3 mode have determined that ESTOP channel 1 cannot be asserted by the internal force-ESTOP circuit. Your channel 1 ESTOP may be wired incorrectly. |
|
-1626 |
ESTOP2 stuck off |
The safety diagnostics in enhanced CAT3 mode have determined that ESTOP channel 2 cannot be asserted by the internal force-ESTOP circuit. Your channel 2 ESTOP may be wired incorrectly. |
|
-1627 |
Motor power stuck on |
The safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is too high when the motor power has been turned off. This may indicate a failed internal safety circuit or an unplugged cable. |
|
-1628 |
Motor power stuck off |
The safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is too low when the motor power has been turned on. This may indicate a failed internal safety circuit or an unplugged cable. |
|
-1629 |
FFC ENA_PWR' signal stuck on |
In a controller that includes a FFC safety board, the safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is too high when the motor power has been turned off. Probably the FFC_ENA_PWR' signal on the FFC board is stuck on or this signal from the controller is shorted high. When the safety board is present, the output from the motor power supply should be off when the controller commands that motor power should be disabled. If the output of motor power supply is on, there is a problem with the motor power disable signal and its associated circuits. |
|
-1630 |
FFC ENA_PWR' signal stuck off |
In a controller that includes a FFC safety board, the safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is too low when the motor power has been turned on. Probably the FFC_ENA_PWR' signal on the FFC board is stuck off or this signal from the controller is shorted low. |
|
-1631 |
Power dump circuit failed |
The safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is not falling quickly enough when power is turned off. This indicates that the power dump circuit has failed or is missing. Verify that you are using the correct robot *.pac files with this robot. |
|
-1632 |
At least one motor must be enabled |
The safety diagnostics in enhanced CAT3 mode have determined that there are no motors enabled for the robot. At least one motor must be present to perform the safety tests. Check the values for DataID 2105 ("Simulate servo interface") and DataID 2026 ("Motor disable mask") to make sure at least one motor is enabled. If you really want all motors disabled, you must also disable safety mode using DataID 117. |
|
-1633 |
Hardware watchdog timer failed to disable power |
The safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is too high after the hardware watchdog timer has expired. There may be a hardware problem with the watchdog timer, or you have attempted to enable safety mode on an older controller that does not contain a hardware watchdog timer. |
|
-1634 |
FPGA watchdog timer failed to disable power |
The safety diagnostics in enhanced CAT3 mode have determined the DC voltage indicated by DataID 12684 ("Nominal DC bus voltage, volt") is too high after the watchdog timer in the FPGA has expired. There may be a hardware problem with the watchdog timer. |
|
-1635 |
FPGA watchdog trigger stuck on |
The FPGA indicates that the watchdog timer is triggered even though it is being polled normally. There may be a hardware problem with the watchdog timer or the FPGA. |
|
-1636 |
FPGA watchdog trigger stuck off |
The FPGA indicates that the watchdog timer is not triggered even though it is not being polled. There may be a hardware problem with the watchdog timer or the FPGA. |